Fig. 3. (A) Sagittal view of the initial CT scan showing that the aortic arch and descending aorta are acutely angulated, leading to an obvious bird-beak formation (white arrow). (B) CT scan reconstruction, posterior view. The CT scan shows distortion of the configuration of the bare spring of the previously implanted endograft (white arrow). (C) CT scan reconstruction, superior oblique view. This view leads to a strong suspicion of fracture of the bare spring of the previously deployed endograft (white arrow). CT, computed tomography.
Vasc Spec Int 2019;35:39-43 https://doi.org/10.5758/vsi.2019.35.1.39